dc.descriptionPoster zaprezentowany na miedzynarodowej konferencji naukowej 1st International Conference on Dosimetry and its Applications, Prague, Czech Republic, 23-28 June 2013; w formie artykułu, po recenzjach został przyjęty do druku w Radiation Physics and Chemistry (2014); DOI: 10.1016/j.radphyschem.2014.05.045pl
dc.description.abstractThe paper reports on three levels of dose rates obtainable from single beta source: (133±3) mGy/s, (17.8±0.3) mGy/s and (1.94±0.04) mGy/s, as calibrated for quartz sand grains. These values were achieved for different attenuation stages of beta radiation emitted by standard Sr-90/Y-90 source with the nominal activity of 1.48 GBq attached to an automatic luminescence reader. Lower dose rates give opportunity for exact dosing, which is especially required in luminescence dating applied to young samples as well as in environmental dosimetry. Moreover new method for determining time lag in opening the source in the Riso beta irradiator is presented. This allowed to resolve the contradiction appearing in the literature. The time delay was found to be (0.15±0.01) s per single irradiation. For improving accuracy the dose rate correction is suggested to be taken into account for irradiations shorter than 30 seconds.pl
dc.description.sponsorshipNCNpl
dc.identifier.citationRadiation Physics and Chemistry, DOI information: 10.1016/j.radphyschem.2014.05.045pl
dc.identifier.issn0969-806X
dc.identifier.urihttp://repozytorium.umk.pl/handle/item/1913
dc.language.isoengpl
dc.publisherElsevier B.V.pl
dc.rightsAttribution-ShareAlike 3.0 Poland*
dc.rightsinfo:eu-repo/semantics/openAccess
dc.rights.urihttp://creativecommons.org/licenses/by-sa/3.0/pl/*
dc.subjectLuminescence dosimetry systempl
dc.subjectDose rate calibrationpl
dc.subjectRetrospective dosimetrypl
dc.subjectBeta radiation sourcepl
dc.titleOn application of low doses from beta radiation source in OSL retrospective dosimetrypl
dc.typeinfo:eu-repo/semantics/articlepl

Files

Original bundle

Loading...
Thumbnail Image
Name:
On application of low doses from beta radiation source in OSL retrospective dosimetry_Poster.pdf
Size:
261.87 KB
Format:
Adobe Portable Document Format

License bundle

Loading...
Thumbnail Image
Name:
license.txt
Size:
1.34 KB
Format:
Item-specific license agreed upon to submission
Description: